QSO3 Inline sensor
Engineered for High UV Intensity
The QSO3 Online Sensor has been developed for reliable UV measurement of medium-pressure UV lamps or UV LEDs in continuous operation. Both the high UV intensity and elevated temperatures present significant challenges for stable measurements. By combining carefully selected materials with a specialized quartz rod system, the QSO3 offers a robust yet easy-to-use solution for online UV monitoring.
Since measurement values are strongly influenced by the mounting position, the QSO3 is designed as a relative measuring device. This means that readings are set to 100% under reference conditions with a new lamp, reflector or comparable setup. The sensor can be conveniently installed close to the UV source, such as at the edges of the reflector, ensuring consistent and practical UV monitoring in real-world applications.
Technical Information
| Configuration | Options |
|---|---|
| UV Measuring range |
UVA 315nm - 395nm UVB 265nm - 325nm UVC 230nm - 280nm UVV 380nm - 445nm UV LED 380nm - 445nm |
| Measuring range |
20 to 2,000 mW/cm² 250 to 25,000 mW/cm² |
| Amplification factors | 1, 10, 100, 1000 and in combination 0.3 |
| Output signal | 0-10 V DC (propotional to irradiance), max. 2 mA |
| Auxiliary voltage | 24 V DC, max. 5 mA |
| Measuring distance to UV source | approx. 20 .... 100 mm |
| Max. permissable intensity |
10,000 mW/cm² for max. 10 min 30,000 mW/cm² for max. 10 s |
| Air purging | 1/8” (max. 0,5 bar) |
| Dimensions (WxHxT) |
approx. 45 x 45 x 34 [mm] stainless steel protection tube: L = 140 mm, ø 8 mm |
| Weight | approx. 200 g |
| Temperature range |
Sensor box: 15°C...50°C (continuously) Quartz rod at measuring window: approx. 300° C End of stainless steel tube: max. 300 °C (continuously) |