QWO Online sensor
Engineered for High UV Intensity
The QWO Online Sensor is specifically engineered for reliable UV medium-pressure lamp monitoring under demanding conditions. Its robust design ensures long-term stability even at high UV intensity and elevated temperatures, where conventional sensors typically drift or degrade. With a proportional output signal of DC 0–10 V, the sensor can be seamlessly integrated into machine control systems. Sensitivity is adjustable up to a factor of 1000 and can be fine-tuned in steps of 0.3, offering flexible adaptation to various application requirements. Easy to handle, durable and precise, the QWO Online Sensor provides dependable UV measurement performance for continuous operation.
Technical Information
| Configuration | Options |
|---|---|
| UV Measuring range |
UVA 315nm - 395nm UVB 265nm - 325nm UVC 230nm - 280nm UVV 380nm - 445nm UV LED 380nm - 445nm |
| Measuring range |
20 to 20,000 mW/cm² 250 to 15,000 mW/cm² |
| Amplification factors | 1, 10, 100, 1000 and in combination 0.3 |
| Output signal | 0-10 V DC (propotional to irradiance), max. 2 mA |
| Auxiliary voltage | 24 V DC, max. 5 mA |
| Measuring distance to UV source | approx. 20 .... 100 mm |
| Max. permissable intensity |
5,000 mW/cm² for max. 10 min 15,000 mW/cm² for max. 10 s |
| Dimensions (WxHxT) | approx. 45 x 45 x 34 [mm] |
| Weight | approx. 135 g |
| Temperature range | 15°C ... 50°C (continuously) |

Contact And Support